High-resolution, high-sensitive ac calorimeter.
View/ Open
Date
1992Author
Bednarz, Grzegorz
Millier, Brian
White, Mary Anne
Metadata
Show full item recordAbstract
A high-resoln., high-sensitivity, automated ac calorimeter capable of heat capacity measurements
of very small samples (m < 20 mg) with a temp. resoln. of a few mK in the ac mode is described.
This calorimeter also can be operated in the relaxation mode to provide abs. heat capacity values
with precision and accuracy of around 2%. This calorimeter was tested by measuring the heat capacity
of gadolinium over its ferromagnetic phase transition. The results agree well with the literature
data giving, however, substantially better resoln. of the heat capacity in the crit. region.
Citation