Automated, low-temperature dielectric relaxation apparatus for measurement of air-sensitive, corrosive, hygroscopic, powdered samples
Date
1999-07
Authors
Bessonette, PWR
White, MA
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
An automated apparatus for dielectric determinations on solid samples was designed to allow
cryogenic measurements on air-sensitive, corrosive, hygroscopic, powdered samples, without
determination of sample thickness, provided that it is uniform. A three-terminal design enabled
measurements that were not affected by errors due to dimensional changes of the sample or the
electrodes with changes in temperature. Meaningful dielectric data could be taken over the frequency
range from 20 Hz to 1 MHz and the temperature range from 12 to 360 K. Tests with Teflon and with
powdered NH4Cl gave results that were accurate within a few percent when compared with literature
values.
Description
Keywords
Citation
Bessonette, PWR, and MA White. 1999. "Automated, low-temperature dielectric relaxation apparatus for measurement of air-sensitive,
corrosive, hygroscopic, powdered samples." Review of Scientific Instruments 70(7): 3113-3114.