Now showing items 1-2 of 2

  • Cavity ringdown spectroscopy measurements of the infrared water vapor continuum 

    Reproduced from Cormier, John G., Roman Ciurylo, and James R. Drummond. 2002. "Cavity ringdown spectroscopy measurements of the infrared water vapor continuum." Journal of Chemical Physics 116(3): 1030-1034, with the permission of AIP Publishing.
    The measurements of the water vapor continuum were analyzed using infrared cavity ringdown spectroscopy (CRDS) at frequencies of 931.002, 944.195 and 969.104 cm-1. The experiments uses a continuous-wave CO2 laser and an ...
  • Infrared water vapor continuum absorption at atmospheric temperatures 

    Reproduced from Cormier, John G., Joseph T. Hodges, and James R. Drummond. 2005. "Infrared water vapor continuum absorption at atmospheric temperatures." Journal of Chemical Physics 122(11), with the permission of AIP Publishing.
    We have used a continuous-wave carbon dioxide laser in a single-mode realization of cavity ring-down spectroscopy to measure absorption coefficients of water vapor at 944 cm-1 for several temperatures in the range 270-315 ...